Anesthesia Machine HART-B
|
|
|
|
Anesthesia Machine HART-B
|
|

-
Domestic classical versatile anesthesia machine, with many working modes like IPPV, SIMV, SIGH. MANUAL, the clinical application scope is very large
-
5" SNT screen to show P-t, V-t dual channel waveform, the technical parameters are obvious to see
-
"Dual electric valve inhalation & exhalation control unit", assuring high precision tidal volume output
HART – B anesthesia machine specifications
| Anesthesia Machine |
|
Power supply: 220V±10%. 5OHz±1Hz Internal battery: DC12V (over for four hours) Power consumption: <50W |
Patient type: Adult, Pediatric
Driven mode: Pneumatically driven electronically controlled Mode: Semi-open. Semi-closed or closed system |
|
|
| Anesthesia Ventilator |
|
Respiration Pattern: IPPV, SIMV. SIGH. MANUAL. Display: 5.5" SNT screen
Waveform: P-t, V-t
Tidal volume: 0. 50ml-1600ml Respiratory frequency: 1 to 65 BPM I: E ratio: 8:1-1:8
Pressure high alarm limit: 20-60cmH20 Pressure low alarm Limit: 0-20cmH20
Monitor Parameters: tidal volume. MV, Ppeak. PEEP, respiratory frequency, I:E ratio. |
Alarm parameters: pressure high alarm pressure low alarm O2 lost alarm
AC power lost alarm Alarm silence function
Safety valve pressure limit: 6Kpa
Good air tightness, high reliability, and minor respiration impede ance |
|
|
| Main Unit |
|
Circle absorber: compact circle absorber Bellow: 0-1600ml
Flow meter: O2,N20 four-tubes flow meter O2 tube: 0L/min-1L/min··10L/min N2Otube: 0L/min-1L/min-10L/min
O2 and N20 Linkage Device
Adjust O2 and N2O in proportion(O2 concentration is not lower than25%) |
Oxygen FLUSH: 35L/min-7SUmin Gas source pressure:
O2: 0.27MPa-O.55MPa
N20: 0.27MPa-O.55MPa
Gauge: two gauges
Anesthetic vaporizer: one vaporizer, two vaporizer is option |
|
|
![]()
|